Description
Technical Description – LaserInspect™ CIMO4
Features
- All in one control panel for Laser Gas Family
- Sleek full colour touchscreen
- 15 inch with narrow tablet style bezels
A Closer Look
- Sleek full-colour touchscreen
- 15” with narrow tablet-style bezels
- Expandable with up to 4 analysers connected to one interface
- Span / Validation reading.
- Touch screen with easy and intuitive navigation
- Easy way to check analyser status
- All-in-one control panel for the LaserGas™ family
- Rugged design for industrial use
- Remote monitoring of instruments
- Control, Inspect, Monitor and keep Overview of up to 4 analysers

The Specs at a Glance
Power Input
- Input voltage: 10 to 30 VDC
- Typical power HMI only: 18 W
- Maximum power HMI only: 24 W
- Maximum power HMI with modules: 58 W
- Must use a Class 2 circuit according to National Electrical Code (NEC), NFPA-70 or Canadian Electrical Code (CEC), Part I, C22.1 or a Limited Power Supply (LPS) according to IEC 60950-1 or Limited-energy circuit according to IEC 61010-1
- Power connection via removable three position terminal block.
Power Connection
- High compression cage-clamp terminal block
- Wire Strip Length: 0.3” (7.5 mm)
- Wire Gauge Capacity: One 14 AWG (1.63 mm) solid, two 18 AWG (1.02 mm) or four 20 AWG (0.81 mm)
Network Media
- 10BaseT: ≥ Cat3 cable
- 100BaseTX: ≥ Cat5 cable
Environmental
- Operating Temperature: -20°C to 60°C
- Storage Temperature: -20°C to 70°C
- Operating Humidity: 0 to 85% (non-condensing)
- Operating Altitude: Up to 2000 metres
- Shock: 40 g per IEC 68-2-27
- Vibration: 4 g @ 5-500 Hz per IEC 68-2-6 LCD Display
LCD Display
- Touchscreen: Resistive analog
- Size: 15”
- Pixels: 1024×768
- Brightness: 400 cd/m2
- Backlight: 70,000 HR TYP